Small field dose delivery evaluations using cone beam optical computed tomography-based polymer gel dosimetry.

Journal of medical physics ·Vol. 36 ·No. 1 ·2011-01-00 ·Pages 3-14

Olding T, Holmes O, Dejean P, McAuley KB, Nkongchu K, Santyr G, Schreiner LJ

Abstract

This paper explores the combination of cone beam optical computed tomography with an N-isopropylacrylamide (NIPAM)-based polymer gel dosimeter for three-dimensional dose imaging of small field deliveries. Initial investigations indicate that cone beam optical imaging of polymer gels is complicated by scattered stray light perturbation. This can lead to significant dosimetry failures in comparison to dose readout by magnetic resonance imaging (MRI). For example, only 60% of the voxels from an optical CT dose readout of a 1 l dosimeter passed a two-dimensional Low's gamma test (at a 3%, 3 mm criteria, relative to a treatment plan for a well-characterized pencil beam delivery). When the same dosimeter was probed by MRI, a 93% pass rate was observed. The optical dose measurement was improved after modifications to the dosimeter preparation, matching its performance with the imaging capabilities of the scanner. With the new dosimeter preparation, 99.7% of the optical CT voxels passed a Low's gamma test at the 3%, 3 mm criteria and 92.7% at a 2%, 2 mm criteria. The fitted interjar dose responses of a small sample set of modified dosimeters prepared (a) from the same gel batch and (b) from different gel batches prepared on the same day were found to be in agreement to within 3.6% and 3.8%, respectively, over the full dose range. Without drawing any statistical conclusions, this experiment gives a preliminary indication that intrabatch or interbatch NIPAM dosimeters prepared on the same day should be suitable for dose sensitivity calibration.

Keywords
Cone beam optical computed tomography polymer gel dosimetry small field dose delivery three dimensional
Authors & Affiliations
7 authors, click to expand affiliations / ORCID
Olding Timothy
Department of Physics, Queen's University, Kingston, Ontario, Canada.
Holmes Oliver
Dejean Paul
McAuley Kim B
Nkongchu Ken
Santyr Giles
Schreiner L John
References (29)
29 references, click to expand
  1. Dose resolution in radiotherapy polymer gel dosimetry: effect of echo spacing in MRI pulse sequence.
    Phys Med Biol. 2001 Feb;46(2):449-60 PMID: 11229725
  2. The storage and manipulation of radiation dose data in a small digital computer.
    Br J Radiol. 1974 Feb;47(554):115-21 PMID: 4206210
  3. Focusing optics of a parallel beam CCD optical tomography apparatus for 3D radiation gel dosimetry.
    Phys Med Biol. 2006 Apr 21;51(8):2055-75 PMID: 16585845
  4. NMR relaxation enhancement in gels polymerized and cross-linked by ionizing radiation: a new approach to 3D dosimetry by MRI.
    Magn Reson Imaging. 1993;11(2):253-8 PMID: 8455435
  5. Initial investigation of a novel light-scattering gel phantom for evaluation of optical CT scanners for radiotherapy gel dosimetry.
    Phys Med Biol. 2007 May 21;52(10):2893-903 PMID: 17473358
  6. Measurement of radiation dose distributions by nuclear magnetic resonance (NMR) imaging.
    Phys Med Biol. 1984 Oct;29(10):1189-97 PMID: 6494247
  7. MR imaging of absorbed dose distributions for radiotherapy using ferrous sulphate gels.
    Phys Med Biol. 1990 Dec;35(12):1623-31 PMID: 2284333
  8. Multiple-spin-echo imaging with a 2D Fourier method.
    Magn Reson Med. 1986 Oct;3(5):707-21 PMID: 3784888
  9. CCD imaging for optical tomography of gel radiation dosimeters.
    Med Phys. 1999 Nov;26(11):2508-13 PMID: 10587241
  10. Three-dimensional dosimetry of small megavoltage radiation fields using radiochromic gels and optical CT scanning.
    Phys Med Biol. 2009 Apr 21;54(8):2463-81 PMID: 19336848
  11. A technique for accurate magnetic resonance imaging in the presence of field inhomogeneities.
    IEEE Trans Med Imaging. 1992;11(3):319-29 PMID: 18222873
  12. Optical-CT gel-dosimetry. II: Optical artifacts and geometrical distortion.
    Med Phys. 2004 May;31(5):1093-104 PMID: 15191297
  13. Evaluation of polymer gels and MRI as a 3-D dosimeter for intensity-modulated radiation therapy.
    Med Phys. 1999 Aug;26(8):1542-51 PMID: 10501054
  14. Evaluation of the gamma dose distribution comparison method.
    Med Phys. 2003 Sep;30(9):2455-64 PMID: 14528967
  15. Polymer gel dosimeters with enhanced sensitivity for use in x-ray CT polymer gel dosimetry.
    Phys Med Biol. 2010 Sep 21;55(18):5269-81 PMID: 20720284
  16. Dose resolution optimization of polymer gel dosimeters using different monomers.
    Phys Med Biol. 2001 Oct;46(10):2665-80 PMID: 11686281
  17. Radiation dose distributions in three dimensions from tomographic optical density scanning of polymer gels: I. Development of an optical scanner.
    Phys Med Biol. 1996 Dec;41(12):2695-704 PMID: 8971963
  18. Polymer gel dosimetry using x-ray computed tomography: a feasibility study.
    Phys Med Biol. 2000 Sep;45(9):2559-71 PMID: 11008956
  19. Three-dimensional dose verification for intensity-modulated radiation therapy in the radiological physics centre head-and-neck phantom using optical computed tomography scans of ferrous xylenol-orange gel dosimeters.
    Int J Radiat Oncol Biol Phys. 2008 Mar 15;70(4):1281-91 PMID: 18313533
  20. A CCD-based optical CT scanner for high-resolution 3D imaging of radiation dose distributions: equipment specifications, optical simulations and preliminary results.
    Phys Med Biol. 2001 Dec;46(12):3191-213 PMID: 11768500
  21. Artefacts in multi-echo T2 imaging for high-precision gel dosimetry: III. Effects of temperature drift during scanning.
    Phys Med Biol. 2001 Oct;46(10):2697-711 PMID: 11686283
  22. Characterization of a parallel-beam CCD optical-CT apparatus for 3D radiation dosimetry.
    Phys Med Biol. 2007 Jul 7;52(13):3693-713 PMID: 17664571
  23. Polymer gel dosimeters with reduced toxicity: a preliminary investigation of the NMR and optical dose-response using different monomers.
    Phys Med Biol. 2006 Jul 21;51(14):3301-14 PMID: 16825731
  24. Radiation dose distributions in three dimensions from tomographic optical density scanning of polymer gels: II. Optical properties of the BANG polymer gel.
    Phys Med Biol. 1996 Dec;41(12):2705-17 PMID: 8971964
  25. The influence of cooling rate on the accuracy of normoxic polymer gel dosimeters.
    Phys Med Biol. 2007 May 21;52(10):2719-28 PMID: 17473347
  26. Cone beam optical computed tomography for gel dosimetry I: scanner characterization.
    Phys Med Biol. 2010 May 21;55(10):2819-40 PMID: 20413828
  27. Optical-CT gel-dosimetry I: basic investigations.
    Med Phys. 2003 Apr;30(4):623-34 PMID: 12722814
  28. Experimental determination of the diffusion coefficient in two-dimensions in ferrous sulphate gels using the finite element method.
    Australas Phys Eng Sci Med. 2001 Mar;24(1):19-30 PMID: 11458569
  29. Optimization of multiple spin-echo sequences for 3D polymer gel dosimetry.
    Phys Med Biol. 2002 Sep 7;47(17):3117-41 PMID: 12361214
Article Info
Journal
Journal of medical physics
Abbr.
J Med Phys
ISSN
1998-3913
Published
2011-01-00
Pages
3-14
Language
English
Region
India
NLM ID
9441104
PMCID
PMC3048952
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: product@genelibs.com