Evaluation of the gamma dose distribution comparison method.

Medical physics ·Vol. 30 ·No. 9 ·2003-09-00 ·Pages 2455-64

Low DA, Dempsey JF

Abstract

The gamma tool was developed to quantitatively compare dose distributions, either measured or calculated. Before computing gamma, the dose and distance scales of the two distributions, referred to as evaluated and reference, are renormalized by dose and distance criteria, respectively. The renormalization allows the dose distribution comparison to be conducted simultaneously along dose and distance axes. The gamma quantity, calculated independently for each reference point, is the minimum distance in the renormalized multidimensional space between the evaluated distribution and the reference point. The gamma quantity degenerates to the dose-difference and distance-to-agreement tests in shallow and very steep dose gradient regions, respectively. Since being introduced, the gamma quantity has been used by investigators to evaluate dose calculation algorithms, and compare dosimetry measurements. This manuscript examines the gamma distribution behavior in two dimensions and evaluates the gamma distribution in the presence of data noise. Noise in the evaluated distribution causes the gamma distribution to be underestimated relative to the no-noise, condition. Noise in the reference distribution adds noise in the gamma distribution in proportion to the normalized dose noise. In typical clinical use, the fraction of points that exceed 3% and 3 mm can be extensive, so we typically use 5% and 2-3 mm in clinical evaluations. For clinical cases, the calculation time is typically 5 minutes for a 1 x 1 mm2 interpolated resolution on an 800 MHz Pentium 4 for a 14.1 x 15.2 cm2 radiographic film.

MeSH Terms
Algorithms Gamma Rays Humans Quality Control Radiometry/methods Radiotherapy Dosage Radiotherapy Planning, Computer-Assisted/methods Radiotherapy, Conformal/methods Reproducibility of Results Sensitivity and Specificity Software Stochastic Processes
Authors & Affiliations
2 authors, click to expand affiliations / ORCID
Low Daniel A
Department of Radiation Oncology, Washington University School of Medicine, St. Louis, Missouri 63110, USA. low@radonc.wustl.edu
Dempsey James F
Article Info
Journal
Medical physics
Abbr.
Med Phys
ISSN
0094-2405
Published
2003-09-00
Pages
2455-64
Language
English
Region
United States
NLM ID
0425746
Subset
IM
Grants
PHS HHS · R01 84409 · United States
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: product@genelibs.com