Screening of urine cultures by three automated systems.

Journal of clinical microbiology ·Vol. 15 ·No. 3 ·1982-03-00 ·Pages 468-74

Pezzlo MT, Tan GL, Peterson EM, de la Maza LM

Abstract

A study was conducted to compare three automated systems and the Gram stain for their ability to detect significant bacteriuria. A total of 1,000 urine specimens were evaluated by Autobac MTS (General Diagnostics), Auto Microbic system (AMS; Vitek Systems, Inc.), and MS-2 (Abbott Laboratories) and compared with a semiquantitative culture plate method. Two hundred thirty-nine (23.9%) specimens had colony counts of >10(5) colony-forming units (CFU)/ml by the culture plate method (group I). Of these, 204 (85.3%) were positive by Autobac, 198 (82.8%) were positive by AMS, and 179 (74.9%) were positive by MS-2. When pure cultures of diphtheroids, lactobacilli, and viridans streptococci not group D were considered contaminants and therefore excluded, there were 118 specimens containing pure cultures of probable pathogens. The percentage of significant isolates detected was 97.4% (115 of 118) by the Gram stain, 96.6% (114 of 118) by Autobac, and 95.8% (113 of 118) by AMS and MS-2. The average detection time for all organisms was 2.2 h by Autobac, 6.1 h by AMS, and 1.8 h by MS-2; therefore, all three methods were more rapid than the 18- to 24-h standard plate culture method. One hundred sixty-one (16.1%) specimens had colony counts of 10(4) to 10(5) CFU/ml (group II). The probable pathogens not detected in this group were two (1.2%) by Autobac and MS-2 and three by AMS (1.9%). The average detection time for group II was 4.2 h by Autobac, 8.9 h by AMS, and 3.8 h by MS-2. Six hundred specimens had colony counts of <10(4) CFU/ml. Of these, 188 had colony counts equal to 10(3) and <10(4) CFU/ml (group III), and 412 cultures were below detectable limits by the standard plate method (group IV). Less than 37 and 15% of groups III and IV, respectively, were detected by instrumentation. Average detection times for groups III and IV were 4.6 and 4.8 h by Autobac, 10 and 11 h by AMS, and 4.2 and 4.4 h by MS-2. The cost of supplies and technical time with Gram stain, Autobac, and MS-2, when used as screening methods, were comparable and considerably less expensive than for the reference method. The AMS was the least expensive system when the cost for identifying probable pathogens was included.

MeSH Terms
Bacteria/isolation & purification Bacteriuria/diagnosis Costs and Cost Analysis Culture Media Humans Urine/microbiology
Chemicals
Culture Media
Authors & Affiliations
4 authors, click to expand affiliations / ORCID
Pezzlo M T
Tan G L
Peterson E M
de la Maza L M
References (13)
13 references, click to expand
  1. Comparative evaluation of the Limulus assay and the direct Gram stain for detection of significant bacteriuria.
    Am J Clin Pathol. 1975 Jan;63(1):142-8 PMID: 46129
  2. Evaluation of an adenosine 5'-triphosphate assay as a screening method to detect significant bacteriuria.
    J Clin Microbiol. 1976 Jan;3(1):42-6 PMID: 767357
  3. Electrochemical method for the early detection of urinary-tract infections.
    Am J Clin Pathol. 1976 Jul;66(1):91-5 PMID: 937276
  4. Microscopy of stained urine smears to determine the need for quantitative culture.
    J Clin Microbiol. 1976 Oct;4(4):372-4 PMID: 61972
  5. Automated microbiological detection/identification system.
    J Clin Microbiol. 1977 Oct;6(4):406-13 PMID: 334798
  6. Impedimetric screening for bacteriuria.
    J Clin Microbiol. 1978 Mar;7(3):273-8 PMID: 348719
  7. Problems of spectrum and bias in evaluating the efficacy of diagnostic tests.
    N Engl J Med. 1978 Oct 26;299(17):926-30 PMID: 692598
  8. Clinical laboratory evaluation of automated microbial detection/identification system in analysis of clinical urine specimens.
    J Clin Microbiol. 1979 Aug;10(2):226-30 PMID: 389950
  9. The Automicrobic System for urines.
    J Clin Microbiol. 1979 Dec;10(6):823-33 PMID: 118176
  10. Rapid semiautomated screening and processing of urine specimens.
    J Clin Microbiol. 1980 Mar;11(3):220-5 PMID: 6991523
  11. Rapid screening for bacteriuria by light scatter photometry (Autobac): a collaborative study.
    J Clin Microbiol. 1981 Jan;13(1):147-50 PMID: 7007418
  12. Evaluation and optimization of urine screening by Autobac.
    J Clin Microbiol. 1981 Apr;13(4):677-80 PMID: 7014617
  13. Asymptomatic infections of the urinary tract.
    Trans Assoc Am Physicians. 1956;69:56-64 PMID: 13380946
Article Info
Journal
Journal of clinical microbiology
Abbr.
J Clin Microbiol
ISSN
0095-1137
Published
1982-03-00
Pages
468-74
Language
English
Region
United States
NLM ID
7505564
PMCID
PMC272119
Subset
IM
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: product@genelibs.com