Home LiteratureArticle Details
PMID: 7008860 Published · ppublish English Journal Article

An improved selected ion recording system for precise isotope ratio determination.

Biomedical mass spectrometry ·Vol. 7 ·No. 9 ·1980-09-00 ·Pages 381-4

Gruenke LD, Craig JC, Bier DM

Abstract

An improved analog four-channel selected ion recording system is described, in which major modifications permit a decrease in the dwell time to 33 ms per channel, thus minimizing the mass cycling error. Synchronization with a sinusoidal sweep voltage superimposed on the normal accelerating voltage (8 kV) enables two channels to be monitored simultaneously in real time and each mass to be recorded continuously. These improvements allow measurement of ion current ratios with a precision of 0.2% over a wide dynamic range, permitting accurate determination of isotopic enrichment in biomedical assays even when this enrichment is derived from a single label. Use of the system is illustrated by the analysis of palmitate turnover in dogs (using [1--13C] palmitic acid) with an average standard deviation corresponding to the detection of 0.04% excess of [1--13C] palmitate.

MeSH Terms
Animals Autoanalysis Dogs Isotope Labeling Mass Spectrometry Palmitates/blood Radioisotope Dilution Technique/instrumentation
Chemicals
Palmitates
Authors & Affiliations
3 authors, click to expand affiliations / ORCID
Gruenke L D
Craig J C
Bier D M
Article Info
Journal
Biomedical mass spectrometry
Abbr.
Biomed Mass Spectrom
ISSN
0306-042X
Published
1980-09-00
Pages
381-4
Language
English
Region
England
NLM ID
0430246
Subset
IM
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: product@genelibs.com