Home LiteratureArticle Details
PMID: 639260 Published · ppublish English Comparative Study Journal Article Research Support, U.S. Gov't, Non-P.H.S. Research Support, U.S. Gov't, P.H.S.

Comparison of double- and single-isotope enzymatic derivative methods for measuring catecholamines in human plasma.

Clinical chemistry ·Vol. 24 ·No. 4 ·1978-04-00 ·Pages 567-70

Evans MI, Halter JB, Porte D

Abstract

We directly compared the reliability of a single-isotope enzymatic derivative technique for measurement of plasma catecholamines with that of the well-established double-isotope method. A significant (p less than 0.001) correlation was observed between measurements (n = 52) in the two assays, both for norepinephrine (r = 0.97) and epinephrine (r = 0.80). Means and coefficients of variation for the two analytes in a pooled specimen of plasma, measured repeatedly during six months, were virtually identical by each assay method. Basal plasma catecholamine concentrations in two different groups of apparently healthy subjects were also similar by each method. Dopamine concentrations in plasma were consistently below the limits measurable by either technique. The single-isotope assay requires half the assay time and 1/200th the sample as the double-isotope method. We conclude that this assay is just as reliable as the double-isotope technique and gives virtually identical values for norepinephrine and epinephrine concentrations in the physiological range.

MeSH Terms
Adult Carbon Radioisotopes Catecholamines/blood Dopamine/blood Epinephrine/blood Evaluation Studies as Topic Humans Methods Middle Aged Norepinephrine/blood Tritium
Chemicals
Carbon Radioisotopes Catecholamines Tritium Dopamine Norepinephrine Epinephrine
Authors & Affiliations
3 authors, click to expand affiliations / ORCID
Evans M I
Halter J B
Porte D
Article Info
Journal
Clinical chemistry
Abbr.
Clin Chem
ISSN
0009-9147
Published
1978-04-00
Pages
567-70
Language
English
Region
England
NLM ID
9421549
Subset
IM
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: product@genelibs.com