Home LiteratureArticle Details
PMID: 6270629 Published · ppublish English Journal Article Research Support, Non-U.S. Gov't

Improved patch-clamp techniques for high-resolution current recording from cells and cell-free membrane patches.

Pflugers Archiv : European journal of physiology ·Vol. 391 ·No. 2 ·1981-08-00 ·Pages 85-100

Hamill OP, Marty A, Neher E, Sakmann B, Sigworth FJ

Abstract

1. The extracellular patch clamp method, which first allowed the detection of single channel currents in biological membranes, has been further refined to enable higher current resolution, direct membrane patch potential control, and physical isolation of membrane patches. 2. A description of a convenient method for the fabrication of patch recording pipettes is given together with procedures followed to achieve giga-seals i.e. pipette-membrane seals with resistances of 10(9) - 10(11) omega. 3. The basic patch clamp recording circuit, and designs for improved frequency response are described along with the present limitations in recording the currents from single channels. 4. Procedures for preparation and recording from three representative cell types are given. Some properties of single acetylcholine-activated channels in muscle membrane are described to illustrate the improved current and time resolution achieved with giga-seals. 5. A description is given of the various ways that patches of membrane can be physically isolated from cells. This isolation enables the recording of single channel currents with well-defined solutions on both sides of the membrane. Two types of isolated cell-free patch configurations can be formed: an inside-out patch with its cytoplasmic membrane face exposed to the bath solution, and an outside-out patch with its extracellular membrane face exposed to the bath solution. 6. The application of the method for the recording of ionic currents and internal dialysis of small cells is considered. Single channel resolution can be achieved when recording from whole cells, if the cell diameter is small (less than 20 micrometer). 7. The wide range of cell types amenable to giga-seal formation is discussed.

MeSH Terms
Acetylcholine/pharmacology Animals Cattle Electrophysiology/instrumentation Fetus/physiology Ion Channels/physiology Membranes/physiology Muscles/cytology,embryology Rana temporaria Rats Receptors, Cholinergic/physiology Time Factors
Chemicals
Ion Channels Receptors, Cholinergic Acetylcholine
Authors & Affiliations
5 authors, click to expand affiliations / ORCID
Hamill O P
Marty A
Neher E
Sakmann B
Sigworth F J
References (18)
18 references, click to expand
  1. Action potentials in the rat chromaffin cell and effects of acetylcholine.
    J Physiol. 1976 Dec;263(3):417-39 PMID: 1018274
  2. Single acetylcholine-activated channels show burst-kinetics in presence of desensitizing concentrations of agonist.
    Nature. 1980 Jul 3;286(5768):71-3 PMID: 6248795
  3. Single Na+ channel currents observed in cultured rat muscle cells.
    Nature. 1980 Oct 2;287(5781):447-9 PMID: 6253802
  4. Properties of internally perfused, voltage-clamped, isolated nerve cell bodies.
    J Gen Physiol. 1978 May;71(5):489-507 PMID: 660159
  5. A receptor for protons in the nerve cell membrane.
    Neuroscience. 1980;5(12):2325-7 PMID: 6970348
  6. Acetylcholine-induced current in perfused rat myoballs.
    J Gen Physiol. 1980 Mar;75(3):297-321 PMID: 7381423
  7. Single-channel currents recorded from membrane of denervated frog muscle fibres.
    Nature. 1976 Apr 29;260(5554):799-802 PMID: 1083489
  8. OVERTURNING AND ANCHORING OF MONOLAYERS.
    Science. 1938 Jun 3;87(2266):493-500 PMID: 17753515
  9. Effect of internal fluoride and phosphate on membrane currents during intracellular dialysis of nerve cells.
    Nature. 1975 Oct 23;257(5528):691-3 PMID: 1186845
  10. Equilibrium dialysis of ions in nystatin-treated red cells.
    Nat New Biol. 1973 Jul 11;244(132):47-9 PMID: 4515993
  11. Effects of permeant monovalent cations on end-plate channels.
    J Physiol. 1979 Mar;288:509-28 PMID: 112241
  12. Single channel currents from excised patches of muscle membrane.
    Proc Natl Acad Sci U S A. 1980 Nov;77(11):6930-4 PMID: 6256772
  13. Inferences about membrane properties from electrical noise measurements.
    Biophys J. 1972 Aug;12(8):1028-47 PMID: 5044577
  14. The extracellular patch clamp: a method for resolving currents through individual open channels in biological membranes.
    Pflugers Arch. 1978 Jul 18;375(2):219-28 PMID: 567789
  15. Separation of sodium and calcium currents in the somatic membrane of mollusc neurones.
    J Physiol. 1977 Sep;270(3):545-68 PMID: 903906
  16. Measured work of deformation and repulsion of lecithin bilayers.
    Proc Natl Acad Sci U S A. 1979 Jun;76(6):2750-4 PMID: 288063
  17. Functional and morphological characterization of isolated bovine adrenal medullary cells.
    J Cell Biol. 1978 Jan;76(1):12-30 PMID: 618889
  18. Shot noise in ion channels.
    Biochim Biophys Acta. 1975 Nov 17;413(1):1-10 PMID: 1191685
Article Info
Journal
Pflugers Archiv : European journal of physiology
Abbr.
Pflugers Arch
ISSN
0031-6768
Published
1981-08-00
Pages
85-100
Language
English
Region
Germany
NLM ID
0154720
Subset
IM
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: product@genelibs.com