Microelectrodes with tip diameters less than or equal to 0.5 micron used for intracellular recording sometimes give artefactual readings due to tip potentials. These appear to be caused by a change in the tip of the microelectrode upon impalement with a corresponding change in microelectrode resistance. A circuit is described which enables the resistance of the microelectrode to be continuously monitored throughout an experiment. The electrode is used as one half of a potential divider for an a.c. signal at 20 Hz with an amplitude of 20 mV. The other half of the potential divider is a 1000 Momega resistor. The a.c. signal across the microelectrode is approximately proportional to the microelectrode resistance in the range 0 --50 Momega. This signal is amplified, rectified and smoothed and then displayed on a chart recorder or other convenient meter.
No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong
Qilu Normal University · Genelibs Bioinformatics Lab
750 Shunhua Rd, Jinan
2F, Bldg F, University Science Park
Tel: 0531-88819269
Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.
Business Email
E-mail: product@genelibs.com