Evaluation of a novel in-line point-of-care blood gas analyser.

Anaesthesia ·Vol. 71 ·No. 9 ·2016-00-00 ·Pages 1044-52

Miles LF, Giraud K, Ferris R, Klein AA, Martinez GC, Jenkins DP, Saulankey K

Abstract

Point-of-care testing is becoming increasingly relevant to the practice of anaesthesia and critical care medicine, especially in terms of minimisation of sample volumes and decreased time to decision making. We performed a prospective observational study to evaluate a novel, in-line blood gas analysis device against a conventional benchtop model, and assessed it while placing the enrolled patients under extreme physiological conditions, specifically deep hypothermic circulatory arrest. Eight patients were studied, and had between seven and 11 samples analysed for seven variables (pH, pCO2 , pO2 , HCO3 (-) , base excess [BE], K(+) and haematocrit [Hct]), using the device during the process of cooling to 20 °C on cardiopulmonary bypass, and subsequent rewarming to normothermia. After Passing-Bablok analysis, the variables were evaluated for bias, limits of agreement and percentage error at above and below 30 °C. Of the measured variables, only pH (percentage error 2.4%) and potassium (19.8%) demonstrated acceptable (< 30%) percentage error over the full range of temperatures measured. Carbon dioxide, when stratified by temperature, was acceptable (< 30 °C percentage error 24.6%, > 30 °C percentage error 9.9%), but the overall percentage error of the dataset (45.8%) was excessively high. Bicarbonate and haematocrit both had an acceptable percentage error above 30 °C (25.2% and 18.5%, respectively), but similar to carbon dioxide, percentage error for the full range of temperatures exceeded 30%. These data differ from previous work examining this device, and highlights the difference between derived measures using different apparatuses when exposed to extreme physiological conditions.

Keywords
ABG values: measured vs. calculated deep hypothermic circulatory arrest hypothermia: pH-stat management point-of-care testing
MeSH Terms
Blood Gas Analysis Cardiopulmonary Bypass Humans Hydrogen-Ion Concentration Point-of-Care Systems Prospective Studies Rewarming
Authors & Affiliations
7 authors, click to expand affiliations / ORCID
Miles L F
Department of Anaesthesia and Intensive Care, Papworth Hospital, Cambridge, UK.
Giraud K
Department of Research and Development, Papworth Hospital, Cambridge, UK.
Ferris R
Department of Cardiothoracic Surgery, Papworth Hospital, Cambridge, UK.
Klein A A
Department of Anaesthesia and Intensive Care, Papworth Hospital, Cambridge, UK.
Martinez G C
Department of Anaesthesia and Intensive Care, Papworth Hospital, Cambridge, UK.
Jenkins D P
Department of Cardiothoracic Surgery, Papworth Hospital, Cambridge, UK.
Saulankey K
Department of Anaesthesia and Intensive Care, Papworth Hospital, Cambridge, UK.
Article Info
Journal
Anaesthesia
Abbr.
Anaesthesia
ISSN
1365-2044
Published
2016-00-00
Pages
1044-52
Language
English
Region
England
NLM ID
0370524
Subset
IM
Corrections
CommentOn
CommentOn
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: product@genelibs.com