Sensitivity of linear CCD array based film scanners used for film dosimetry.

Medical physics ·Vol. 33 ·No. 11 ·2006-11-00 ·Pages 3993-6

Devic S, Wang YZ, Tomic N, Podgorsak EB

Abstract

Film dosimetry is commonly performed by using linear CCD array transmission optical densitometers. However, these devices suffer from a variation in response along the detector array. If not properly corrected for, this nonuniformity may lead to significant overestimations of the measured dose as one approaches regions close to the edges of the scanning region. In this note, we present measurements of the spatial response of an AGFA Arcus II document scanner used for radiochromic film dosimetry. Results and methods presented in this work can be generalized to other CCD based transmission scanners used for film dosimetry employing either radiochromic or radiographic films.

MeSH Terms
Computer Peripherals Equipment Design Equipment Failure Analysis Film Dosimetry/instrumentation,methods Radiation Dosage Reproducibility of Results Sensitivity and Specificity Signal Processing, Computer-Assisted/instrumentation
Authors & Affiliations
4 authors, click to expand affiliations / ORCID
Devic Slobodan
Medical Physics Department, McGill University Health Centre, Montreal, Quebec, Canada. devic@medphys.mcgill.ca
Wang Yi-Zhen
Tomic Nada
Podgorsak Ervin B
Article Info
Journal
Medical physics
Abbr.
Med Phys
ISSN
0094-2405
Published
2006-11-00
Pages
3993-6
Language
English
Region
United States
NLM ID
0425746
Subset
IM
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: product@genelibs.com